GA200 Main Page

GA2000 Infrared Gas Analyser

Specifications
Gas Ranges
Detection Principle CO2 and CH4 by dual wavelength infra-red cell with reference channel.  O2, CO and H2S by internal electrochemical cells
Oxygen cell lifetime approximately 3 years in air
CH4  CO2  O2  CO  H2S 
Range 0-70% to specification, 0-100% of reading  0-40% to specification, 0-100% of reading 0-25% 0-500 ppm 0-200 ppm
Typical Accuracy 
0-5% vol  +0.5% +0.5% +1% - -
5-15% vol +1% +1% +1% - -
15%- FS +3% +3% +3% - -
0-Full scale +10%  +10%
Response Time, T90 <20 seconds <20 seconds <20 seconds <60 seconds <60 seconds
Power Supply
Battery type Rechargeable NmH battery pack. Not user replaceable. LiMn battery for data retention
Battery life Typical use, 10 hours
Alternative power Can be powered externally for fixed in place applications. Call for futher information.
Battery lifetime Up to 1000 charge/discharge cycles
Facilities
Temperature measurement With optional probe, -10 deg C to +75 deg C
Temperature accuracy + 0.2 deg C
Visual and audible alarm User selectable CO2, CH4 and O2 alarm levels via DataField software
Communications  RS232
Relative pressure + 250 mbar from calibration pressure
Pump
Typical Flow 300 cc/min
Flow fail point 50 cc/min approx.
Flow with 200 mbar vacuum 250 cc/min approx.
Vacuum pull 400 mbar approx.
Environmental Conditions
operating temperature range 0 to 40 deg C
Physical
Weight  2 kg
Size Length 630 mm, Width 190 mm, Depth 252 mm
Case material anti static ABS
Display  LCD, 40 x 16 characters, Fiber optic woven back light for low light conditions
Filters  User replaceable integral fiber filter at inlet port and an external PTFE water trap filter

    

 