Size: |
|
10" L x 3.0" W x 2.5" H (25.5 cm x 7.6 cm x
6.4 cm) |
|
Range:
|
|
0 to 999.9 ppm
1000 to 15,000 ppm |
|
Weight: |
|
26 oz (738 g) |
|
Resolution: |
|
0.1 ppm, 1 ppm |
|
Sensors: |
|
Photoionization sensor with standard 10.6 eV or optional 9.8
eV or 11.7 eV lamps
|
|
Response: |
|
3 sec. |
|
|
|
|
Measurement: |
|
Isobutylene +2 ppm
or 10% of val. <2,000ppm |
|
Battery: |
|
Rechargeable, external field-replaceable Lithium-Ion battery
pack
Alkaline battery adapter
|
|
Accuracy: |
|
+/-20% of reading
> 2,000 ppm |
|
|
|
|
PID
Detector: |
|
Easy access to lamp
& sensor |
|
|
|
|
Correction
Factors: |
|
Built-in 102 VOC
gases |
|
Operating
Hrs: |
|
16 hours of operation (12 hours with alkaline
battery) |
|
Calibration: |
|
2 point field cal.
of 0 and std reference gas calibration memory
of 8 calibration curves, alarm limits,
span values and calibration date |
|
Batt.
Charging: |
|
Up to 10 hrs charge
w/built-in smart charger |
|
|
|
|
|
Display: |
|
4 lines, 28 x 43 mm, with LED backlight for enhanced display
readability
|
|
Inlet
Probe: |
|
Robust, Flexible
125mm tubing |
|
Keypads: |
|
1 operation and 2 programming keys, 1
flashlight on/off |
|
Direct
Readout: |
|
Instantaneous reading VOCs as ppm by volume
High values
STEL and TWA
Battery and shutdown voltage
Date, time, temperature |
|
Alarms: |
|
95 dB (at 30 cm) buzzer and flashing red LED to indicate
exceeded preset limits
High: 3 beeps and flashes per second
Low: 2 beeps and flashes per second
STEL and TWA: 1 beep and flash per second
Alarms latching with manual override or automatic reset
Additional diagnostic alarm and display message for low
battery and pump stall |
|
EMI/RFI:: |
|
Highly resistant to EMI/RFI
Compliant with EMC Directive 89/336/EEC |
|
IP Rating: |
|
IP67 unit off and without flexible probe
IP65 unit running |
|
Datalogging: |
|
Standard 6 months at one-minute intervals |
|
Calibration: |
|
Two-point or three-point calibration for zero and span.
Calibration memory for 8 calibration gases, alarm limits,
span values and calibration dates |
|
Sampling Pump: |
|
Internal, integrated flow rate at 500 cc/mn
Sample from 100' (30m) horizontally and vertically |
|
Low Flow Alarm: |
|
Auto pump shutoff at low-flow condition |
|
Communication: |
|
• Download data and upload instrument set-up from PC through
charging cradle or optional Bluetooth™
Wireless data transmission through built-in RF modem |
|
Frequency: |
|
902 to 928 MHz (license-free), 2.400 to 2.4835 GHz
(license-free), 433 MHz, 869 MHz |
|
RF Range: |
|
Up to 500' (900 MHz, 433 Mhz, 869 Mhz) extendable with
RAELink3 Repeater to 2 miles |
|
Hazard area approval: |
|
US and Canada:
UL, cUL, Classified as Intrinsically
Safe for use in Class I, Division I Groups A, B, C,
D
Europe: ATEX II 1G EEx ia IIC T4 (pending)
IECEx: II 1G EEx ia IIC T4 (pending) |
|
Temperature: |
|
-4° to 113° F (-20° to 50° C) |
|
Humidity: |
|
0% to 95% relative humidity (non-condensing) |
|
Attachments: |
|
Durable bright yellow rubber boot with belt
clip |
|
Warranty: |
|
Lifetime on non-consumable components (per RAE Systems
Standard Warranty), 3-year warranty for 10.6 eV lamp,
1 year for pump and batter |
|